32
FEP630, FEH630
ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
… 5 Electrical connections
… Pin assignment
Electrical data for inputs and outputs
Note
• An additional document with Ex safety instructions is
available for measuring systems that are used in potentially
explosive atmospheres.
• Ex safety instructions are an integral part of this manual. As
a result, it is crucial that the installation guidelines and
connection values it lists are also observed.
The icon on the name plate indicates
the following:
Power supply
AC power supply
Terminals
L / N
Operating voltage
100 to 240 V AC (
Æ
15 % / +10 %), 47 to 64 Hz
Power consumption
S
max
: < 20 VA
Power-up current
18.4 A, t < 3 ms
DC voltage supply
Terminals
1+ / 2
Æ
Operating voltage
16.8 to 30 V DC
Ripple
< 5 %
Power consumption
P
max
: < 20 W
Power-up current
21 A, t < 10 ms
Current output Uco / 32, 31 / 32
Can be configured for outputting mass flow and volume flow via
the on-site software.
G11596-02
+
-
A
B
I
E
32-
U
co
32-
I
E
31+
R
B
R
B
U
q
31+
U
co
+
-
A
Current output Uco / 32, active
B
Current output 31 / 32, passive
Figure 41: (I = internal, E = external, RB = load)
G10323-02
0
0
100
200
300
400
500
600
5
10
15
20
25
30
35
RB
[Ω]
Uq [V]
Permissible source voltage U
q
for passive outputs in relation to load
resistance R
B
where I
max
= 22 mA.
JJJ
= Permissible range
Figure 42: Source voltage for passive outputs
Current output
Active
Passive
Terminals
Uco / 32
31 / 32
Output signal
4 to 20 mA or
4 to 12 to 20 mA switchable
4 to 20 mA
Load R
B
250
έ
á
R
B
á
300
έ
250
έ
á
R
B
á
600
έ
Source voltage U
q
* —
13
V
á
U
q
á
30 V
Measuring error
< 0.1 % of measured value
Resolution 0.4
μ
A per digit
Insulation
The current output and digital outputs are electrically
isolated.
* Source
voltage
U
q
depends on the load R
B
and must be within the
permissible range.
For information on communication via the HART protocol, refer
to
HART® Communication
on page 43.