99
FEM630 |
ELECTROMAGNETIC FLOWMETER | OI/FEM630-EN REV. A
Diagnostics / . . .Diagnosis Control / . . .Diagnosis Conductiv*
Conductivity On/Off
Activate the conductivity diagnostic function.
Default setting: Off
Note
Conductivity diagnostic can be used in the nominal diameter range of 3/8” to 12” (DN 10 to 300).
For additional information, see
Extended diagnostic functions
on page 118.
Conductivity[µS/cm]
Indicator of the measured conductivity in μS/cm
Adj. Cond. Value
Conductivity must be set in accordance with the conditions on-site.
Measure the conductivity using a conductivity meter on-site and enter the measured value here.
Limits: 5 to 20000 μS/cm
Cond. Iout Min Value
The conductivity value is available as a 4 to 20 mA-output (option card).
Cond. Iout Max Value
Set the 4 mA and 20 mA value which correspond to the upper and lower range of the conductivity value.
Cond.Min Alarm Value
Set the alarm for minimum and maximum conductivity. In the case of down-scale, an alarm is triggered.
Cond.Max Alarm Value
Limits: 5 to 20000 μS/cm
Elec. Imp. E1-GND
Electrical impedance E1-GND.
Current impedance between electrode E1 and GND (ground potential).
Elec. Imp. E2-GND
Electrical impedance E2-GND.
Current impedance between electrode E2 and GND (ground potential).
Diagnostics / . . .Diagnosis Control / . . .Diagnosis SIL**
SIL On/Off
For information purposes only.
SIL devices are delivered ex works as SIL devices. There is no special SIL mode to activate
8 . . .Operation
. . .Parameter descriptions
*
The menu is only available if the Expert Diagnosis function is activated. See also the ‘Device Setup\...Transmitter\...Feature Settings’ menu.
**
Menu only available if SIL diagnostic function is activated. See also the ‘Device Setup\...Transmitter\...Feature Settings’ menu.