Navigator 600 Phosphate
Multi-stream
4 Installation
18
IM/NAV6P/MS–EN Rev. F
4.6.1 Connections Overview
Note.
Field wiring terminal screws must be tightened to a torque of 0.6 to 0.8 Nm (5 to 7 lbf.in).
Fig. 4.8 Connections Overview
O/P 6
O/P 5
O/P 4
O/P 3
O/P 2
O/P 1
Socket for
Optional Digital
Communications
Ethernet RJ45
Connector
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
Hold/Maintenance
Relay
Calibration Failed
Relay
Alarm 2 Relay
Alarm 4 Relay
Alarm 6 Relay
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
Out of Service
Relay
Calibration in
Progress Relay
Alarm 1 Relay
Alarm 3 Relay
Alarm 5 Relay
AC Fuse 1 A Type T
or
**DC Fuse 12.5 A Type T
100 V to 240 V AC ±10 %
(90 V min. to 264 V max.)
50 / 60 Hz
L
N
E
+
–
E
TB7
Current Outputs
TB8
Optional Digital
Communications
TB2
TB3
TB4
TB
5
TB6
TB1
Use fuse rated:
AC Supply
1 A (max.), Type T, 250 V
DC Supply*
12.5 A 125 V DC Type T
Manufacturer*
SCHURTER,
Model SPT 5 x 20 Series
18 to 36 V DC
DC fuse replacement details