Sensyflow FMT200-ECO2

 THERMAL MASS FLOWMETER  |  OI/FMT200-ECO2-EN REV. H 

15

 

 

Parameter descriptions 

 

 

  

 

  

 

NOTICE 

Data loss 

Data loss due to the shut-off of power supply during the 
initialization phase (status LED blinks red) of the device. 
•  Do not switch off the power supply until the device is 

ready for operation(status LED is continuous green). 

 

 

Saving the parameterization 

If any parameterization has been changed in the 
parameterization program and transferred to the device through 
the 

‘Accept’

 button, a save / initialization phase of the device 

follows and the status LED blinks red. 
•  The supply voltage must not be switched off until the status 

LED is continuous green once more. 

 
The modified parameters will not be saved properly if this rule is 
not obeyed.  
After the repeated power-up of the device, a data inconsistency 
will be detected during the initialization phase, the status LED 
will switch to continuous red and the device will no longer 
function correctly. 
 

Reestablish the parameterization 

Once the status LED is continuous red, communication with the 
parameterization software can be resumed.  
1.  The parameterization data must be resaved by selecting the 

‘Accept’

 button. 

•  Wait until the symbol indicating that a program is running 

disappears and the status indication for the ‘initialization 
phase’ is terminated. 

•  The status LED is still continuous red at this stage. 

2.  Perform a hardware reset by powering down and powering 

up the power supply. 
•  After the initialization phase, the status LED is 

continuous green and the device is ready for operation. 

 

Note 

If operation readiness cannot be reestablished, contact ABB 
Service. 
 

 

 

 

Menu bar 

 

Menu / parameter 

Description 

Configuration 

 

Factory settings 

Shows the parameters set at the factory 

Savings 

Saves the current parameter set to a data storage 
medium. 

Load 

Loads a saved parameter set from a data storage 
medium. 

Delete 

Deletes a saved parameter set from a data storage 
medium. 

Print 

Prints the current parameter set to the configured 
standard printer. 

Select Language 

Select the program language: 
German, English, French 

Select unit 

Select the flow unit from a unit list. 

Insert / delete unit 

Create / delete a new unit. 
Refer to 

Add user-defined unit

 on page 16. 

Exit 

Exits the parameterization program. 

Connection 

 

Connection 

Select the serial interface for connection with the 
sensor. 

Data backup 

 

Data backup 

Save the changed data to the device 

 

 

Summary of Contents for FMT200-ECO2

Page 1: ...00 ECO2 is a very compact and highly dynamic mass flowmeter for air The Sensyflow FMT200 ECO2 enables direct mass or standard volume flow measurement in pneumatic applications for example for the cont...

Page 2: ...ll the sensor 9 Electrical connections 10 Terminal layout 10 Electrical data for inputs and outputs 10 Parameterize 11 6 Commissioning 12 Checks prior to commissioning 12 Switching on the power supply...

Page 3: ...to the product may only be performed if expressly permitted by these instructions Information and symbols on the product must be observed These may not be removed and must be fully legible at all time...

Page 4: ...between the product and your network or any other network as the case may be Operator shall establish and maintain any appropriate measures such as but not limited to the installation of firewalls app...

Page 5: ...bar abs High pressure version 16 102 kPa 16 bar abs Process connections Small flange adapter ISO KF flange screwed connection G in G in G in G 1 in Legris hose adapter Transair adapter Wetted materia...

Page 6: ...m resistor is maintained at a constant overtemperature in relation to an unheated platinum sensor inside the gas flow The heating power required for maintaining the overtemperature depends directly on...

Page 7: ...corrugated cardboard and place it in a box of sufficient size lined with a shock absorbing material e g foam rubber The thickness of the padding should be appropriate for the device weight and type o...

Page 8: ...mounting position can be selected Inlet sections To prevent negative effects on measuring accuracy the ISO KF flange DN 25 Transair G in and G 1 should be provided in the inlet sections when using the...

Page 9: ...the small flange adapter DN 25 decreasing approx quadratically at smaller flow rates Installing the sensor Before installing the sensor observe the following points The ambient conditions at the insta...

Page 10: ...alog output signal 2 Brown RS 232C TxD transmi data 3 Green Pulse frequency output signal 4 Yellow Power supply 24 V DC 5 Gray Power supply GND 6 Pink RS 232C RxD receive data 7 Blue Analog output GND...

Page 11: ...op bits 1 Parity None Data bits 8 Parameterize The Sensyflow FMT200 ECO2 can simultaneously operate one analog output current 0 4 to 20 mA or voltage 0 to 5 10 V one digital output frequency pulse ala...

Page 12: ...anges and output signals in line with requirements 6 Commissioning Checks prior to commissioning The following points must be checked before commissioning the device Correct wiring in accordance with...

Page 13: ...the PC laptop using the D Sub adapter If no RS232 interface is available additionally use the USB RS232 adapter 2 Power up the power supply of the sensor 3 To launch the parameterization program execu...

Page 14: ...rview CURRENT DEVICE DATA ANALOG OUTPUT CURRENT 0 to 20 mA 4 to 20 mA VOLTAGE 0 to 10 V 0 to 5 V DIGITAL OUTPUT FREQUENCY OUTPUT 10 to 1000 Hz 1 to 100 Hz PULSE OUTPUT VARIABLE 1 to 2500 Hz SWITCH OUT...

Page 15: ...umed 1 The parameterization data must be resaved by selecting the Accept button Wait until the symbol indicating that a program is running disappears and the status indication for the initialization p...

Page 16: ...multiple of kg h Parameter Description Unit Name of the unit text field Factor Enter the factor of the unit in kg h Decimal numbers must be entered using a period rather than a comma Standard conditi...

Page 17: ...screen 1 Menu bar 2 Registration Tab 3 Device identification 4 Current measured value 5 Standard condition parameter display when selecting a volume unit 6 Status bar 7 Sensor connection status 8 Ana...

Page 18: ...f the current measured value up scales the set upper limit the bar has a red color If the current measured value is within the set limits the bar has a green color Standard condition Figure 11 Standar...

Page 19: ...Set flow rate per impulse Duration Set impulse length Polarity Set polarity low high for the impulse Switch option Figure 14 Switch The current measured value is monitored for fluctuation either side...

Page 20: ...Parameterization program Analog Output Registration Tab This tab is used to configure the analog output For the sake of transparency both the current and amended values are displayed at the same time...

Page 21: ...he maximum current maximum voltage on the analog output Low flow Set the flow rate below which no measurement is made measured value 0 Filter factor Set the time constant for the damping of the output...

Page 22: ...meterization program Digital Output Registration Tab This tab is used to configure the digital output For the sake of transparency both the current and amended values are displayed at the same time Th...

Page 23: ...quency corresponds to a flow rate of 100 of the set upper range value Parameter Description min Enter the frequency for 0 flow rate max Enter the frequency for 100 flow rate maximum 2500 Hz Impulse ou...

Page 24: ...status LED on the sensor Status LED on the transmitter The status LED on the transmitter indicates the operating condition of the device and displays possible faults LED Description Continuous green l...

Page 25: ...nor should they be bent or damaged in any other way Any attempts to clean the sensor that go beyond the steps described above may cause destruction to the sensor Cleaning through blow out 1 Remove the...

Page 26: ...sideration If the device is to be used at another location the device should preferably be packaged in its original packing so that it cannot be damaged Observe the notices in Returning devices on pag...

Page 27: ...Sensyflow FMT200 ECO2 THERMAL MASS FLOWMETER OI FMT200 ECO2 EN REV H 27 11 Accessories Sensor Figure 21 Flowmeter sensor FMT200 ECO2 with mounted small flange adapter...

Page 28: ...Process connection KF DN 25 to adapt inlet run and outlet run includes 2 clamp rings and 2 sealing rings Inlet section 10 D both sides with KF DN 25 connection Outlet section 5 D both sides with KF DN...

Page 29: ...ter pair for inlet and outlet inlet adapter with high tech flow conditioner Legris hose adapter pair Screwed connection G in simultaneous connection for Transair system 25 mm pair Transair adapter 25...

Page 30: ...ification Note The device data sheet is available in the ABB download area at www abb com flow 13 Additional documents Note All documentation declarations of conformity and certificates are available...

Page 31: ...ils Company Address Contact person Telephone Fax Email Device details Type Serial no Reason for the return description of the defect Was this device used in conjunction with substances which pose a th...

Page 32: ...otice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We res...

Reviews: