Test gas configurations – AutoCal systems with restrictors
System using air as test gas 1 and bottled gas as
(optional) test gas 2
Figure 25 Test gas supply, AutoCal system with restrictors –
air (test gas 1) and optional bottled gas (test gas 2)
System using 2 bottled test gases
Figure 26 Test gas supply, AutoCal system with restrictors –
2 bottled test gases
AZ10 probe
AZ10 AutoCal unit
Test gas supply: restrictor in Autocal
unit limits flow to 2.2 l/min (4.662 scfh)
at 1 bar (15 psi)
Test gas 1/1 bar (15 psi)
Test gas port
5 μm/oil
coalescing
filter regulator
Test gas 2 (optional) from bottled gas –
adjusted to 1 bar (15 psi) output
Clean dry oil-free instrument air
INPUT
: 10 bar (145 psi) max.
OUTPUT
: 1 bar (15 psi) max
Non-return
valve
TG1
TG2
Test gas port
Non-return
valve
Test gas 2 (optional) from bottled gas –
adjusted to 1 bar (15 psi) output
Test gas 1 from bottled gas –
adjusted to 1 bar (15 psi) output
AZ10 probe
TG1
TG2
Test gas 1/1 bar (15 psi)
AZ10 AutoCal unit
Test gas supply: restrictor in Autocal
unit limits flow to 2.2 l/min (4.662 scfh)
at 1 bar (15 psi)