A Z10 OX YG E N A N A LY ZE R
| CO M B US T I O N G A S A N A LY S I S | O I/A Z 10 - EN R E V. A
6 Installation
General mounting requirements
NOTICE
• For optimum performance, mount the probe at 90° to the
gas flow.
• The probe will function adequately in turbulent gas flows.
• Do not obstruct the flue gas exit hole.
Figure 7 General mounting requirements
X
X
45 º
45º
45 º
Duct surface temperature 400 °C (752 °F) max.
If the flue is lagged, a stand-off tube of
suitable length with a minimum internal
diameter of 55 mm (2.16 in) may be used
Do not mount the probe in such a manner that
allows ash/soot/debris to accumulate in the probe
Ash/soot/debris will
accumulate here and
block the probe cell
Ambient air temperature at
probe terminal head
–20 to 70 °C (–4 to 158 °F)
max.
Exit
Gas flow
See
Notice
above
See
Notice
above
See
Notice
above
Gas flow
Gas flow
Gas flow
Exit
Exit
Exit
55 mm
(2.16 in) ∆
min.
max.
max.