Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
IM/AZ20E–EN Rev. F
29
Signals View
The
Signals View
displays active signals and their values as a list.
Values for signals that have not been configured are not
displayed.
Use the
and
keys to scroll through active signals.
Signals Units
Mnemonic
Displayed
Description
Typical Values
O2
TMP
dT
mV
O/P
CJ
Off
Fct
zOf
Dev
d
RT
RR
Vac
Frq
Int
%O2
°C or °F
°C / Min or °F / Min
mV
%
°C or °F
mV
mV
%O2
K
K
/ Year
s
%O2 / s
V
Hz
°C or °F
% Oxygen
Cell Temperature
Rate of Rise of
Cell Temperature
Cell Millivolts
Heater Output
Cold Junction Temp
Calibration Offset
Calibration Factor
Cell Zero Offset
O
2
Deviation
Cell Impedance
Cell Impedance Drift
Test Gas Response Time
Recovery Rate
Mains Voltage
Mains Frequency
Internal Temperature
0.01 to 100.00 % O
2
>700 °C (1292 °F)
After stabilization phase
0 to 60 °C / min
(0 to 108 °F / min)
–50 to 190 mV
0 to 100 %
–20 to 70 °C (–4 to 158 °F)
–20 mV to 20 mV
0.900 to 1.100
<20 mV
<0.5 % O
2
0.1 to 10.0 k
<1 k
8 to 30 s
0.05 to 0.30 % O
2
90 to 265 V
50 to 60 Hz
–20 to 70 °C (–4 to 158 °F)
Display Overview
Section and Function
Table 5.3 Operator Menu Screens (Continued)
Signals View
Exit
O2 xx.x
%O2
TMP
xxx °C
dT
x.x deg C/m
Back
O2 xx.x
%O2