Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
IM/AZ20E–EN Rev. F
57
…Input/Output
…Digital I/O 1 (2)
Type
Selects the type of automatic calibrations started by a
digital input – see Section 6.1, page 64.
Available only if the
Input Function
is a
'Start Autocal',
or 'Start/Stop Autocal'
and the other I/O is not
Calibration Type
.
One point
One point
Two point
Calibration
Sequence
Selects the type of calibration / check performed
when activated by a digital input.
Calibration
Performs a calibration that updates the calibration
offset and factor automatically if successful.
Accuracy Check
Performs an accuracy check (the normal calibration
sequence) but does not update the calibration offset
and factor in use by the transmitter.
Cal. On Error Limit
Performs an accuracy check but, if the error is greater
than the value set in the
Cal. Error Limit
field (see
page 41), the calibration offset and factor are
updated automatically.
Warn On Error Limit
Performs an accuracy check, but if the error is greater
than the value set in the
Cal. Error Limit
field (see
page 41) a calibration warning is generated.
Parameter
Comment / [Range]
Default
Table 5.12 Input/Output Menu (Continued)