Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
IM/AZ20E–EN Rev. F
47
…Device Setup
Security Setup
Used to set
Standard
and
Advanced
level passwords of
up to 6 alphanumeric characters.
Note
.
Standard
and
Advanced
passwords are not set at
the factory and must be added by the end-user(s).
None
Standard
Set by
Standard
and
Advanced
users.
Advanced
Set by
Advanced
users.
Reset Fact. Settings
Resets all configuration parameters to their default values.
Factory-set
Reset Temp. Trip
For heated probes (AZ20 and AZ30), if the cell
temperature exceeds a preset limit (approx. 850 °C
[1562 °F]), a hardware over- temperature trip automatically
activates and switches off power to the heater.
If the cell temperature remains at this elevated
temperature for less than 30 minutes, the trip is reset
automatically.
If the elevated temperature remains for more than 30
minutes, it is necessary to reset the trip (at this menu).
Alternatively, the trip can be reset by switching power to
the transmitter off and on again.
N / A
Parameter
Comment / [Range]
Default
Table 5.9 Device Setup Menu (Continued)