Endura ACA592
Toroidal conductivity transmitter
8 Troubleshooting and Diagnostics
70
OI/ACA592/TC–EN Rev. B
8.2.3 ACA592 Toroidal Conductivity Transmitter – Diagnosis Simulation
All diagnosis conditions except
Loop test
(see page 64) can be simulated individually using the DTM or
EDD.
To simulate a diagnosis, turn diagnosis simulation ON, select a diagnosis and send it to the transmitter. In
Simulated Diagnosis
mode, all active diagnoses are suppressed and only the simulated diagnosis is
reported.
Diagnosis simulation status can be viewed on the HMI using
Diagnostics / Diagnosis status
Section 7.3.6, page 60). The status is either
Simulated
or
Actual
(diagnosis simulation cannot be set or
cleared from the HMI.
Caution.
Diagnosis simulation cannot be performed via the transmitter's display.
Diagnosis simulation should be used for advanced test purposes only.
Diagnosis simulation is volatile – simulation setting is lost after a power-up.