Endura ACA592
Toroidal conductivity transmitter
7 Operator Pages and Menus
60
OI/ACA592/TC–EN Rev. B
7.3.6 Diagnostics
Used to perform a loop test (output verification) and indicate
the diagnostic status (actual / simulated).
Diagnostics Level
Parameter
Description
Loop Test
The current output of the transmitter can be adjusted between 3.8 and 21.5 mA.
This feature is useful to verify that the transmitter output matches the current in the control
room.
Enabled when
HART Poll Addr = 0
– see page 61.
Diagnosis Status
Used to determine if diagnostics simulation is active (read-only).
The display should read 'Actual' indicating real-time diagnostics from the transmitter.
6
123
Menu
Exit
Select
Diagnostics