Endura ACA592
Toroidal conductivity transmitter
7 Operator Pages and Menus
58
OI/ACA592/TC–EN Rev. B
7.3.4 Input/Output
Used to setup process value high / low ranges, damping and
alarm current settings.
Input Output Level
Parameter
Description
PV Rng Units
Enables conductivity units selection for the purpose of editing the PV range values (
PV Lo /
Hi Rng Val
). This is independent of the Process value units. Parameter is read-only when
Analyzer Type
=
Concentration
.
PV Lo Rng Val
The PV low (minimum) value 0.0 (corresponding to 4 mA) of selected
PV Rng Units
.
PV Hi Rng Val
The PV high (maximum) value corresponding to 20 mA of selected
PV Rng Units
.
Maximum value:
2,000,000 µS/cm (or 2000 mS/cm)
Damping
Enables damping of the 4 to 20 mA response.
Minimum value 0.0 s, maximum value 99.9 s.
Alarm Current
Level of current loop output under alarm conditions.
Options comprise:
High
(corresponds to alarm current level of 21.5 mA)
Low
(corresponds to alarm current level of 3.8 mA)