![ABB Endura ACA592-TC Operating Instructions Manual Download Page 54](http://html1.mh-extra.com/html/abb/endura-aca592-tc/endura-aca592-tc_operating-instructions-manual_3052232054.webp)
Endura ACA592
Toroidal conductivity transmitter
7 Operator Pages and Menus
52
OI/ACA592/TC–EN Rev. B
7.3 ACA592 Toroidal Conductivity Transmitter Parameters
7.3.1 Easy Setup
Users with
Advanced
level access have read and write
privileges to the parameters listed.
Users with
Standard
and
Read Only
access have read-only
privileges to parameters at
Easy Setup
level.
Easy Setup Level
Parameter
Description
Analyzer Type
Selects the measurement type:
Conductivity
Concentration
Conc. Soln
.
Enabled if
Analyzer Type = Concentration.
Sets the concentration solution from the following options:
0–18% HCI, 0–20% H
2
SO
4
, User Def., 0–15% NaOH or 0–20% NaC1.
Conc. Unit
Enabled if
Analyzer Type = Concentration
.
Selects the concentration unit:
%, ppm, µg/l, ppb, mg/l, none
Conc. Curve Name
Enabled if
Analyzer Type = Concentration
.
A user-defined name.
Press the
key to enter the required characters for the name as detailed on page 40.
Temp. unit
(Temperature Units)
Sets the temperature unit (°C or °F) applied to all transmitter temperature parameters and
values.
Temp. Comp. Type
(Temperature Compensation Type)
Manual
Manual
temperature compensation is enabled when the device does not detect any
temperature sensor connection either at power-up or upon execution of
Run TC
Recognition
. When
Temp Comp Type
is
Manual
, the user configurable
Manual Temp
value is
used for temperature compensation.
Auto
Enabled when a temperature sensor is detected by the device either at power-up or when
executing the
Run TC Recognition
(see page 56) routine.
1
Menu
Exit
Select
Easy Setup