Endura ACA592
Toroidal conductivity transmitter
8 Troubleshooting and Diagnostics
OI/ACA592/TC–EN Rev. B
65
F088.016
Sensor electronics failure
or wrong sensor board
[Sensor electronics
interface error]
Data exchange with sensor
electronics board has failed.
[ELECTRONICS]
Power cycle the device and check if the problem is
corrected.
If the error persists, replace electronics.
F106.032
Unreliable output current
The D to A converter is not properly
Calibrated / Trimmed or output
current has reached its hardware
limit.
[ELECTRONICS]
Perform an Output Trimming.
If the error persists, replace the electronics.
Check the device configuration.
F108.035
Output ReadBack failure
The output circuit could be broken
or not correctly calibrated
[ELECTRONICS]
Perform an Output Trimming routine.
If the error persists replace electronics.
Diagnostic Message
[Sub-error Message]
Possible Cause(s)
[Related Condition]
Corrective Action
Table 8.2 Diagnostic Messages (Continued)