![ABB CGWB 13 Operator'S Manual Download Page 31](http://html.mh-extra.com/html/abb/cgwb-13/cgwb-13_operators-manual_4010031.webp)
29
TBL 01-C
Self-regulating heating conductor
TBL 01-E
Fixed resistor heating conductor
Outer covering
Metal corrugated tubing (for ACK) or
Outer sheath of PVC (for ACMP)
Polyamide corrugated tubing
Sample gas hose
PTFE tube 6 x 4 x 1 mm
PTFE tube 6 x 4 x 1 mm
or 8 x 6 x 1 mm
End assembly of
tubing
Probe: metal cap - analyzer: silicone cap
(mandatory for ACK) or with POM plastic
caps on each side (mandatory for ACMP) or
probe: PTFE cap - analyzer: POM cap
(mandatory for ACMP 120 °C)
PA hard caps
Power supply
230 V or 115 V, 50…60 Hz
230 V or 115 V, 50…60 Hz
Heating power
30 W/m or 47 W/m
90 W/m
Holding temperature 100 °C or 120 °C
200 °C
Ambient temperature –30 … +60 °C
–30 … +60 °C
Outer diameter
43 mm
43 mm
Maximum line length 70 m
40 m
Line bushing
M48
M48
Minimum bending
radius
200 mm
200 mm
Weight
approx. 1 kg/m
approx. 1 kg/m
Electrical connection
for heater and
Pt100 100
permanently connected, 3 x 2.5 mm
2
or
2 x 0.14 mm
2
, length 5 m
permanently connected, 3 x 2.5 mm
2
or
2 x 0.14 mm
2
, length 5 m