![ABB CGWB 13 Operator'S Manual Download Page 29](http://html.mh-extra.com/html/abb/cgwb-13/cgwb-13_operators-manual_4010029.webp)
27
Heatable Sample Gas Line CGWB 13
Line with frost protection
Self-regulating strip heater Parallel strip heater,
regulated
Outer
covering
PVC PVC PVC
Sample gas hose
PTFE
PTFE
PTFE
Line end assembly
various options
(see specification sheet)
various options
(see specification sheet)
various options
(see specification sheet)
Heating
self-regulating strip heater
self-regulating strip heater
CPD heating element,
regulated
Power supply
230 V or 115 V, 50…60 Hz
230 V or 115 V, 50…60 Hz
230 V or 115 V, 50…60 Hz
Heating power
16 W/m
30 W/m
60 W/m
Current consumption Making/hold current
= 65/25 mA
0.15 A/m
0.3 A/m
Holding temperature +20…+50 °C
+90…+110 °C
200 °C
Ambient temperature –30…+60 °C
–30…+60 °C
–30…+60 °C
Outer diameter
30 mm
43 mm
40 mm
Sample gas hose
6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm
Maximum line length 130 m
90 m
60 m
Line
bushing
M42 M42 M42
Minimum bending
radius
300 mm
300 mm
300 mm
Weight
approx. 1 kg/m
approx. 1 kg/m
approx. 1 kg/m
Connecting cable for
heating (permanently
connected)
3 x 2.5 mm
2
; length 5 m
3 x 2.5 mm
2
; length 5 m
3 x 2.5 mm
2
; length 5 m
Connecting cable for
Pt100 (permanently
connected)
–
–
LIYCI 2 x 0.14 mm
2
;
length 5 m
C
H A P T E R
4
Technical Data