39
5
PROGRAMMING…
…5.6
Configure Outputs
Zero Value
pH
(or
mV
or
Deg.C
or
Deg.F
or
A-B
) and
Adjust
are shown alternately on the center display
line. Use the
and
keys to adjust the displayed reading to the required zero value:
Note.
The zero value setting plus the minimum differential determines the minimum and
maximum values for the span setting, e.g. to set a span value of –1100mV, first set the zero
value to –1200mV.
pH
– –2.00 to 14.00 pH (minimum differential, 2.00pH)
ORP/Redox
– –1200 to 1100 mV (minimum differential, 100mV)
Temperature
– Deg. C –10.0 to 140.0 (minimum differential, 10
C)
Deg. F 14.0 to 284.0 (minimum differential, 18
F)
A-B
– –2.00 to 12.00 pH (minimum differential, 2.00pH)
Note.
A-B
is applicable only to dual input analyzers.
Default Output
Select the system reaction to failure:
Hold
– Hold the analog output at the value prior to the failure.
On
– Stop on failure. This drives the analog output to the level set in the
Default Val
frame below.
Off
– Ignore failure and continue operation.
Default Value
The level to which the analog output is driven if a failure occurs.
Set the value between 0.00 and 22.00mA
Output 2 configuration (and Outputs 3 and 4 if option board fitted
and
analog
features enabled – see Section 7.3) is identical to Output 1 configuration.
Option board fitted
and
analog features enabled (Section 7.3) – see Section 5.7.
Option board fitted
and
Serial Communications feature enabled (Section 7.3) –
see Supplementary Manual
PROFIBUS Datalink Description (IM/PROBUS).
Single input analyzer
option board not fitted – see
PID Control
Supplement
(
IM/AX4PID)
.
Dual input analyzer
option board not fitted – see Section 5.8.
AO1: Zero Value
7.00
pH
0.0
pH
AO1: Default
-----
mA
Hold
On
Off
AO1: Default Val
12.00
mA
Config. Output 2
Config. Output 1
Config. Output 1
On
Off
or
Hold
CONFIG. SERIAL
CONFIG. CLOCK
CONFIG. SECURITY
CONFIG. CONTROL