21
Solution A: Enter point 1 (to 5)
Deg.C
and
Adjust
are shown alternately on the upper display line. Using the
and
keys, adjust the temperature reading (in 5
increments) to the first of the temperatures on
the pH/temperature curve.
pH
and
Adjust
are shown alternately on the center display line. Using the
and
keys,
adjust the pH reading (in 0.01pH increments) to the pH reading that corresponds to the
temperature reading entered above.
Notes.
1) For accurate calibration, it is important to repeat the above for buffer solution A at
all
5
points along the pH/temperature curve.
2) The displayed temperature value increases automatically by 5
C from the value set for
the previous point. The setting may be increased but not decreased.
Solution B: Enter point 1 (to 5)
Solution B set up is identical to solution A set up.
Note.
For accurate calibration, it is important to repeat the above for buffer solution B at
all
5 points along the pH/temperature curve.
The analyzer calculates the pH/temperature relationship from the data entered.
See Section 4.1.3.
Alter Sec. Code
not set to zero (Section 5.8) – see Section 5.1.
Alter Sec. Code
set to zero (Section 5.8) – see Section 5.2.
…4.1
Sensor Calibration
4.1.2
Set Up User Defined Buffers (pH Only)
4
SETUP…
A: Enter point 5
35.0
Deg.C
4.02
pH
Sensor Cal. A
SECURITY CODE
Set Auto Buffers
Buffer Type
set to
User
(see Section 4.1.1)
A: Enter point 1
15.0
Deg.C
0.00
pH
B: Enter point 5
35.0
Deg.C
4.02
pH
B: Enter point 1
15.0
Deg.C
3.99
pH
CONFIG. DISPLAY
A: Enter point 1
15.0
Deg.C
3.99
pH
Summary of Contents for AX416
Page 70: ...68 NOTES ...