PROGRAMMING
5.5
Configure Alarms
Configure Alarm 1
Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted
and
analog features enabled – see Section 7.3) is identical to Alarm 1
configuration.
Alarm 3 can also be configured as a Wash alarm if
A3: Type
is set to
Wash
–
see following frame.
Alarm 1 Type
Select the type of alarm required:
Off
– The alarm is disabled, the alarm LED is off and the relay is de-energized
at all times.
Alarm
– The analyzer is configured using the
Assign
frame (following) to
generate an alarm in response to a specified high or low pH,
Redox (ORP) or process temperature sensor reading.
Status
– An alarm is generated if either a power failure or a condition that causes
any of the error messages in Table 8.1 (page 62) to be displayed
occurs.
Wash
– Alarm 3 is configured to control the wash sequence.
Note.
The
Wash
alarm type can be assigned only to Alarm 3 and is displayed only when
the lower display line shows
A3: Type
.
A1: Type
set to
Off
or
Status
.
A1: Type
set to
Alarm
– continued below.
A3: Type
set to
Wash
– see Section 5.5.1.
Alarm 1 Assign
Select the alarm assignment required:
Sen.A
– The analyzer activates an alarm if the pH or Redox (ORP) value of the process
Sen.B
fluid measured by the selected sensor exceeds or drops below the value set
in the
Alarm 1 Set Point
parameter, depending on the type of
Alarm 1 Action
selected – see next page.
Temp.A
– The analyzer activates an alarm if the temperature of the process fluid
Temp.B
measured by the selected sensor exceeds or drops below the value set in
the
Alarm 1 Set Point
parameter, depending on the type of
Alarm 1 Action
selected – see next page.
A–B
– The analyzer activates an alarm if the difference between the Sensor A and
Sensor B readings exceeds or drops below the value set in the
Alarm 1 Set
Point
parameter, depending on the type of
Alarm 1 Action
selected – see
next page.
Note.
The
Sen.B
,
Temp.B
and
A–B
alarm assignment types are applicable only to dual
input analyzers and
A–B
is displayed only when
Probe Type
for each sensor is set to
pH
–
see Section 5.3.
Continued on next page.
Config. Alarm 1
-----
A1: Type
-----
A1: Assign
-----
CONFIG. ALARMS
-----
Wash
Status
Alarm
Off
A-B
Temp.B
Sen.B
Temp.A
Sen.A
A1: Failsafe
Config. Alarm 2
A1: Assign
Config. Alarm 1
Wash Mode
Summary of Contents for AX416