42/23-572 EN Rev. 5
Operator’s Manual ACF-NT, ACF-NT V0309
29
Cleaning/Replacing the Sample Gas Filter in the SC Block,
continued
Step Action
Remove the sample gas filter assembly from the sample conditioning
block and disassemble it.
3
1
2
4
5
6
5
7
8
3
4
Clean the sample gas filter
6
in an ultrasonic bath. Use an aqueous
cleaner (e.g. Extran). Wash the filter element several times with
distilled water and acetone and dry it thoroughly. If a VOC analyzer is
integrated in the analyzer system all hydrocarbons must be removed
to avoid drift of the VOC analyzer.
Replace the O-ring
3
in the sample gas filter cover
2
.
Replace the O-ring on the back of the bottom plate
4
(not visible in
the photo) and the two PTFE profile gaskets
5
.
5
Always use new O-rings and gaskets with a cleaned sample
gas filter. Contaminated or damaged O-rings and gaskets will
reduce sample gas path seal integrity and lead to erroneous
measurement values.
6
Re-assemble the sample gas filter assembly and place it in the sample
conditioning block.
7
Place sample gas filter cover
2
on the sample conditioning block and
secure it with three mounting screws
1
. Tighten mounting screws only
sufficiently to achieve metal-to-metal contact of the sample gas filter
holder.