![ABB ACF-NT Operator'S Manual Download Page 26](http://html.mh-extra.com/html/abb/acf-nt/acf-nt_operators-manual_4305026.webp)
26
Operator’s Manual ACF-NT, ACF-NT V0309
42/23-572 EN Rev. 5
Replacing the Filter Element in the Probe,
continued
Step Action
1
Turn the T-handle
1
of the filter
removal device
1-3
in counter-
clockwise direction.
This pulls the filter element
8
via the
detaching disk
3
out of the casing
10
.
2 Turn
bridge
2
until it can be pulled off
from the bridge holding device
9
through the elongated holes.
3
Pull out filter element
8
with bridge
2
and detaching disk
3
.
4
Turn detaching disk
3
until it can be
pulled off from the hexagon screws
5
via the elongated holes.
Replacing the
Filter Element
Never loosen or tighten the hexagon screws
5
. They have been
adjusted at the factory so that the detaching disk
3
can be
easily moved.