Wear Part Replacement
CAUTION!
Prior to performing any maintenance works on the analyzer system be sure
to activate the Maintenance Control on the “Control Panel” screen (see
page 14) thus setting the “Maintenance Mode” status signal.
Also set the Flow Control on the “Control Panel” screen either to “Zero gas
local” or “Zero gas probe” to avoid any contact to the measuring gas.
Be sure to reset these settings after finishing the maintenance work.
Component Part
No.
Replacement
Instructions
Gas sampling probe filter
or filter stone
0730683
0730682
every 3 months
see Page 25
SC block filter
0768914
every 6 months
see Page 28
Air purifier inlet filter
0999755
every 6 months
see Page 30
Replacing Wear Parts
Air conditioning unit filter mesh
0999765
every 3 months
see Page 30
Use of FFKM Probe
Seals in the System
Version for Measure-
ment of HF
The filter on the gas sampling probe has FFKM seals.
When they are replaced (see section “Replacing the Filter Element in the Probe”,
page 25), FFKM seals must also be used. For quick replacement, it is recom-
mended that a complete, preassembled filter element be kept at the ready. The
following wearing parts are required:
Component Part
No.
Comments
Filter stone
730682
The two FFKM O-ring seals from part
number 801994 must be used.
FFKM seal set
801994
Two O-ring seals on the filter stone
FFKM seal set
730722
Filter fixing device with three O-ring seals
Battery on the
System Controller
Type “Sonnenschein SL-360/S” or “Saft LS14500”
Spare Parts
Information
Spare parts information can be found on the Internet in the “Spare Parts Infor-
mation and Ordering System Parts OnLine” using the address
http://www.abb.com/partsonline.