24
…8.3
Factory Settings Page
Calibrate Temperature Zero
Set the temperature simulator to 100
Ω
and allow the instrument display to stabilize.
Advance to next parameter.
Calibrate Temperature Span
Set the temperature simulator to 150
Ω
and allow the instrument display to stabilize.
Advance to next parameter.
Adjust Retransmission 1 Zero
Set the milliammeter reading to 4mA.
✶
Note.
The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 1 Span
Set the milliammeter reading to 20mA.
✶
Note.
The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Zero
(available only on 4623/800 and 4628/800 instruments)
Set the milliammeter reading to 4mA.
✶
Note.
The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Span
(available only on 4623/800 and 4628/800 instruments)
Set the milliammeter reading to 20mA.
✶
Note.
The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Alter Factory Code
Set
the factory settings access code to a value between 00000 and 19999.
Return to
Operating Page.
…8
CALIBRATION
Temp Zero (100R)
x x x x x
Temp Span (150R)
x x x x x
Adjust RTX Zero
– – – – –
Adjust RTX Span
– – – – –
Adjust RTX Zero2
– – – – –
Adjust RTX Span2
– – – – –
microsiemens/cm
5 0 0
Return to
top of page
Alter Fact. Code
0 0 0 0 0