16
1.30
µ
S/cm xx.x
°
C
Alarm 2 Setpoint
1.00
Alarm 1 Setpoint
µ
S/cm at 25.0
°
C
Operating Page
Section 6.2 , Page 15
Set Up
Alarms Page
Section 7.4, Page 18
Factory Settings Page
Section 8.3, Page 23
Operating Parameters
4.00
4.00
0.50
Access to Secure Parameters
Section 7.1, Page 17
00000
SECURITY CODE
Secure Parameters
Language Page
Section 7.2, Page 17
Espanol
Francais
Deutsch
English
–––––
0.05
SET UP PARAMETER
–––––
Set Up Parameters Page
Section 7.3, Page 17
10.00
Display Span
Cell Constant
–––––
Temp Units (
°
C)
SET UP ALARMS
A1 Action High
Low
A1 Setpoint
–––––
–––––
A1 Type Fail
Temp
–––––
Cond
Off
A2 Action High
Low
A2 Setpoint
–––––
1.00
A2 Type USP
Temp
–––––
Cond
Off
Alter Sec
.
Code
00000
Fail
or
Of
f
Fail
or
Of
f
Fail
USP Alarm Offset
0.20
Cond
or
T
emp
USP
Set Up Retransmission Page
Section 7.5, Page 20
–––––
–––––
SET UP RETRANS
Set Up Retrans 1
RTX Type 4-20
RTX Span
µ
S/cm
–––––
–––––
–––––
100.0
0-20
0-10
0-20
0-10
Set Up Retrans 2
RTX Type 4-20
Cond
RTX O/P Temp
–––––
0.0
Test Retrans (%)
0.0
RTX Zero
µ
S/cm
100.0
Cond
RTX Span
µ
S/cm
No
–––––
–––––
Calibrate Yes
–––––
No
Ye
s
xxxxx
xxxxx
xxxxx
xxxxx
xxxxx
xxxxx
FACTORY SETTINGS
–––––
00000
00000
Alter Fact
.
Code
Adjust RTX Span
Adjust RTX Zero
Temp Span (150R)
Temp Zero (100R)
Res Span 2 (1k0)
Res Zero 2 (O/C)
Res Span 1 (10k)
Res Zero 1 (O/C)
FACTORY SET CODE
ELECTRICAL CAL
–––––
Adjust RTX Span2
Adjust RTX Zero2
–––––
–––––
Operating Parameters
Secure Parameters
A
vailable only on 4623/800
and 4628/800 instruments
7
PROGRAMMING
Fig. 7.1 Overall Programming Chart
✶
Note.
All parameter values shown on the
upper display are the Company standard
settings.
Summary of Contents for 4623
Page 30: ...28 NOTES ...