21
7
PROGRAMMING
…7.5
Set Up Retransmission Page
Set Up Retransmission 2
Note.
Available only on 4623/800 and 4628/800 instruments.
Retransmission 2 Output Range
Set the retransmission output current range for retransmission channel 2.
Retransmission 2 Output Assignment
Select the Retransmission output required:
Temp
– Temperature
Cond
– Conductivity
Retransmission 2 Span
Set the required span over which the retransmission output is to operate to between 10
and 100% of the display span – see Section 7.3,
Set Up Parameters Page
.
Retransmission 2 Zero
Set the zero point for the retransmission 2 output.
Test Retransmission Output
The instrument automatically transmits a test signal of 0, 25, 50, 75 or 100% of the
retransmission range selected above. The % test signal selected is shown on the upper
display.
Example
– for a selected range of 0 to 20mA and 50% retransmission test signal, 10mA
is transmitted.
Select the required retransmission test signal.
Advance to
Factory Settings Page
.
Set Up Retrans 2
– – – – –
RTX Type 4-20
0-20
0-10
– – – – –
Cond
– – – – –
RTX O/P Temp
RTX Span
µ
S/cm
RTX Zero
µ
S/cm
Test Retrans (%)
Temp
Cond
FACTORY SETTINGS
– – – – –
Return to
top of page
0 . 0
0 . 0
1 0 0 . 0
Summary of Contents for 4623