3 Installation
20 |
266 MODBUS® multivariable transmitter User manual 2105216 Rev. AB
Vertical flange transmitter direct-mount instructions
To direct-mount the assembly vertically:
1.
Determine where the transmitter is to be placed on the meter run so that the high side (+) is upstream (Figure 14).
266 multivariable transmitter
RTD
Meter run
Manifold
Figure 14: Vertical flange assembly, direct mount (for illustrative purposes only)
2.
Align the sealing rings and bolt holes, then bolt the transmitter to the manifold.
3.
Screw the risers into the flange taps on the meter run.
4.
Attach the manifold and transmitter assembly to the risers and secure.
The transmitter is now direct-mounted.
Summary of Contents for 266 MODBUS
Page 163: ......