1.3 Specifications
Power Requirements
See Section 1.2 Model Number Breakdown
Power Consumption
Refer to signal converter instruction manual for
power input requirements; typically, 30VA.
Flowmeter Characteristics
Meter Size/Flow Capacity
See Table 1-1
Span
Factory set at specified range between extremes
listed in Table 1-1; can be field adjusted.
Rangeability
100:1
Minimum Liquid Conductivity
5
µ
S/cm
System Accuracy
Flowrate
Accuracy
Frequency Output
< 2% Cal Factor
≥
2% Cal Factor
±
0.02% Cal Factor
±
0.5% of rate
RFI Protection
Class 2-abc-0.1% (10 V/m-20 to 1000 MHz) Per
SAMA Standard PMC 33.1-1978. Standard with
converters; not available with remote converter.
Environmental Limits
Enclosure Classification with
or without Signal Converter
IEC 529 IP65; NEMA 4X
Accidental Submergence
IE 529 IP67, 30 feet H
2
O/48 h (9 m H
2
O/48 h)
Environmental Temperature Limits
at Flowmeter.
−
40 to 150
o
F
(-40 to 65
o
C
)
Process Limits
Temperature Limits, Process Liquid
May not exceed 356
o
F (180
o
C)
Thermal Shock
positive temp change, 212
o
F (100
o
C)
negative temp change, 158
o
F (70
o
C)
10D1476 K-MAG MAGNETIC FLOWMETER INSTRUCTION MANUAL
1-6