With the exception of the primary board, the associated electronics package is called the signal
converter and is remotely mounted. Figure 1-2 shows the electronics compartment for the remote
signal converter. The remote mounted signal converter improves performance if 1) the vibration
specification is exceeded and/or 2) the process liquid temperature vs. ambient temperature exceeds
the value specified for that of an integral converter.
The signal converter also contains a magnet driver unit that is used to power the meter’s magnet
coils. The steady state magnetic field principle, referred to as the MAG-X design concept, provides
optimum zero point stability at an optimized frequency which is a submultiple of the power line
frequency.
For additional information, refer to Section 4.0.
1.2 Model Number Breakdown
Refer to the equipment’s data sheet or the data tag on the equipment for the model number of the
instrument furnished. The details of a specific number are as follows:
10D1476 K-MAG MAGNETIC FLOWMETER INSTRUCTION MANUAL
1-3