6.4
Control light to indicate malfunction
Malfunction is indicated by blinking of the red light emitting diode on the
control panel. In case of malfunction, please first turn off the device and
then on again. Should the device still indicate a malfunction, charge the
mega battery pack in case of low voltage.
In cases of malfunction such as the blinking of 1 – 4 times we consciously
abstained from automatically activating the electro-magnetic brakes as
soon as the malfunction is recognized while the device is driving
.
F
If you want to activate the electro-magnetic brake, push the rocker
switch hard to overcome the pressure point and the panic switch is acti-
vated. Then release the rocker switch
.
In all other cases of malfunction the electro-magnetic brake is automati-
cally activated
.
Caution
:
Should the display blink red 7 times (overheating) then one has
to wait for approximately 5 minutes, depending on the temperatur
.
F
In this condition the device cannot be activated by first switching it off
and then on again
.
F
Should you have to move the wheelchair and the device is in said conditi-
on of malfunction, you may only do so by mechanically locking the slide
switches of the power unit (see chapter
3.3) .
F
After the device cooled down it turns itself back on automatically.
For all other cases of malfunction, please contact your dealer.
26
Red diode is blinking
Malfunction
1 x
Motorcontroller, motor oder mega battery pack defective
2 x
low voltage, battery is empty
3 x
overvoltage
4 x
electro-magnetic brake at fault
5 x
indicator for excess voltage
6 x
indicator for overstressing
7 x
indicator for excess temperature