31
8 5 Safety check
i
NOTE - Safety checks may only be performed by the manufacturer or your authorized dealer!
i
NOTE - We recommend a safety check every two years! The inspection plate informs you about
the next date for inspection. We also recommend a safety check, if the CR2303 is transferred
to a new user.
8 6 Change of operator / re-use
A re-use is referred, if the AAT product of patient X is transmitted to patient Y.
The CR2303 is suitable for re-use. We recommend a safety check as well as a cleaning/desinfection, if
the CR2303 goes to a new user. (see chapter 8.3 and 8.5)
i
NOTE - You may not mend or by-pass a damaged fuse!
Always replace it with an identical one of the same amperage (40 A)!
i
NOTE - Should your fuses burn out frequently, please contact AAT Alber Antriebstechnik GmbH
or your authorized dealer.
Remove the locking plate on
the left side of the battery
pack by loosening the screw.
» Use a screwdriver.
Pull out the covers and the
old fuses, then insert the new
fuses and replace the covers.
»
AAT offers you a replace
-
ment of the fuses 40 A.
Close with locking plate.
» Fix it by tightening the
screw with a screwdriver.
8 4 Replacing fuses
Incorrect use may damage the unit and in turn impair its technical functionality.
This is why this safety check should be carried out every 2 years or before each
subsequent use. We also recommend a safety check, if the CR2303 is transfer-
red to a new user.
To perform a safety check special training is required. More detailed information is
available on the website: www.aat-online.de