A more complex mission to monitor a large frequency spectrum with a single SPECTRAN
®
V6 device, using the
Spectran V6 Sweep Zoom and Wrapped Spectrum blocks.
Download this base mission with all settings at
https://v6-forum.aaronia.de/demo-mission-1/
Problems and solutions
If starting the RTSA suite PRO
does not start the measurement
automatically.
the measurement manually by clicking on the Start/Stop button in the ribbon bar should work
without problem.
Pressing the “Connect” button
Check that all cables are connected correctly, and that the power supply is properly plugged into
the power socket.
The status display in the ribbon
error.
The USB connection does not provide enough bandwidth to transfer all data from the device to the
PC. There are two types of causes for this:
▷ The “
DATA” connection does not support USB 3 SuperSpeed. Please make sure the port is con-
nected to a USB 3.x port on the PC or the cable does support the USB 3 “SuperSpeed” standard.
▷ The cur
rent measurement parameters are set too high. Some parameter combinations require
more bandwidth than a single USB 3 connection can handle. This can be caused by using both Rx1
and Rx2 in parallel, or using more than 122 MHz RTBW on a RSA 2000X model. When using such set-
ups please also connect the “BOOST” cable in addition to the “DATA” cable.
The Status display in the ribbon
-
-
ration. Either reduce the amount of data generated (lower Receiver Clock setting or use Spectra
example by reducing the FFT size in the IQ Power Spectrum block settings).
The device appears to be wor-
king (fan is on), but no data is
displayed.
settings. In particular using the “Spectra” connectors while using IQ “Output Format”, or using “IQ 2” or
“Spectra 2” outputs without changing the “Receiver Channels” setting.