background image

A more complex mission to monitor a large frequency spectrum with a single SPECTRAN

®

 V6 device, using the 

Spectran V6 Sweep Zoom and Wrapped Spectrum blocks.

Download this base mission with all settings at 

https://v6-forum.aaronia.de/demo-mission-1/

Problems and solutions

If starting the RTSA suite PRO 

does not start the measurement 

automatically.

the measurement manually by clicking on the Start/Stop button in the ribbon bar should work 

without problem.

Pressing the “Connect” button 

Check that all cables are connected correctly, and that the power supply is properly plugged into 

the power socket.

The status display in the ribbon 

error.

The USB connection does not provide enough bandwidth to transfer all data from the device to the 

PC. There are two types of causes for this:

▷ The “

DATA” connection does not support USB 3 SuperSpeed. Please make sure the port is con-

nected to a USB 3.x port on the PC or the cable does support the USB 3 “SuperSpeed” standard.

▷ The cur

rent measurement parameters are set too high. Some parameter combinations require 

more bandwidth than a single USB 3 connection can handle. This can be caused by using both Rx1 

and Rx2 in parallel, or using more than 122 MHz RTBW on a RSA 2000X model. When using such set-

ups please also connect the “BOOST” cable in addition to the “DATA” cable.

The Status display in the ribbon 

-

-

ration. Either reduce the amount of data generated (lower Receiver Clock setting or use Spectra 

example by reducing the FFT size in the IQ Power Spectrum block settings).

The device appears to be wor-

king (fan is on), but no data is 

displayed.

settings. In particular using the “Spectra” connectors while using IQ “Output Format”, or using “IQ 2” or 

“Spectra 2” outputs without changing the “Receiver Channels” setting.

© AARONIA AG - Subject to modifications and errors. 

Summary of Contents for SPECTRAN V6

Page 1: ...Check package contents Before your new SPECTRAN V6 can be used please verify that all necessary components are available and all requirements are met USB Stick with manual application software and ot...

Page 2: ...USB connection when operating at full capacity In such scenarios connect the black cable labeled Boost to the BOOST port of the device and a second USB 3 x port on your PC Do not use a USB hub for con...

Page 3: ...ment providing signal strengths of up to 20 dBm not suppor ted on RSA 250X RX TX REF One SMB input labeled REF to connect an external 10 MHz clock signal One SMB in output labeled TRG for synchronizin...

Page 4: ...ar within a few seconds Attention Linux users The RTSA also works on Linux but the installer is not included on the USB stick and needs to be downloaded from our website For obvious reasons the instal...

Page 5: ...all In the same way the FFT parameters in the IQ Power Spectrum block can be accessed or the processing or display parameters of any other block Configuring the SPECTRAN V6 RSA The Spectran V6 block p...

Page 6: ...hannels Receiver Select Output Format Receiver Clock Start Stop Connect Disconnect Center Frequency RF Amplifier Frequency Profile Status Display Span Attention Changing these parameters can result in...

Page 7: ...nfigurations One of the simplest missions possible The mission begins with a calibration block on the left in the blockgraph editor see picture above To specify which cable antenna and or other RF equ...

Page 8: ...tion does not support USB 3 SuperSpeed Please make sure the port is con nected to a USB 3 x port on the PC or the cable does support the USB 3 SuperSpeed standard The current measurement parameters ar...