CONNECTORS
WARNING:
Avoid overloading the input by feeding too much power, otherwise the highly
sensitive electronics may be destroyed!
The maximum permitted peak power that can be applied at the input of the
SPECTRAN V5 is:
- Activated Preamplifier: Max. 0dBm (Peak)
- Deactivated Preamplifier: Max. +20dBm (Peak)
2.
Connectors
2.1. Power Cable / Battery Charging
Standardized 3.5mm mono „jack“. The input
voltage must be between 5V and 5.9V DC (di
-
rect current). The included power supply pro-
vides a voltage of 5.5V at up to 6A (33W).
While being operated, the analyzer needs about
23W, without internal battery only 18W. To oper-
ate the device connect the charger to a power
supply, connect the jack plug of the charger with
the power connector of the SPECTRAN. The
charging LED of the SPECTRAN should turn
red.
Now the internal battery is being charged au-
tomatically. A complete charge cycle takes ap-
proximately 8-10 hours. When the battery is fully
charged, the charging LED turns green (power
supply has to be connected).