July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
Page 21
Temperature
Float
1
The temperature in degree centigrade. The value is used for
calculation of Salinity, Sound speed and Density. Can be
altered while sensor is running to compensate for variable
temperature.
3)
DS
No /
Low
Enable Derived
Sound Speed
BOOL 1
If enabled it will calculate the sound speed from the other
available sensors or fixed values. This overrides the Sound
Speed setting. If a parameter is available from another sensor it
will use this in the calculation, otherwise it will use the fixed
value.
Pressure Sensor Id
ENUM 1
When the DCPS is connected to AiCaP, a list of other available
sensors connected to the SmartGuard/SeaGuard Datalogger is
shown. This makes it possible to get a correct depth input from
another sensor through the datalogger
7)
Only in AiCaP mode
Pressure Parameter
Id
ENUM 1
A list of all available Pressure parameters with kPa as unit is
shown in the dropdown menu. Make sure that the correct
pressure parameter is selected. Only in AiCaP mode.
Air Pressure Sensor
Id
ENUM 1
When the DCPS is connected to AiCaP, a list of other available
sensors connected to the SmartGuard/SeaGuard Datalogger is
shown. This makes it possible to get a correct air pressure input
from another sensor through the datalogger
7)
Only in AiCaP
mode
Air Pressure
Parameter Id
ENUM 1
A list of all available Air Pressure parameters with hPa or mbar
as unit is shown in the dropdown menu. Make sure that the
correct pressure parameter is selected. Only in AiCaP mode.
Temperature Sensor
Id
ENUM 1
When the DCPS is connected to AiCaP, a list of other available
sensors connected to the SmartGuard/SeaGuard Datalogger is
shown. This makes it possible to get a correct temperature input
from another sensor through the datalogger
7)
Only in AiCaP
mode
Temperature
Parameter Id
ENUM 1
A list of all available Temperature parameters with DegC or
Deg.C as unit is shown in the dropdown menu. Make sure that
the correct temperature parameter is selected. Only in AiCaP
mode.
Conductivity Sensor
Id
ENUM 1
When the DCPS is connected to AiCaP, a list of other available
sensors connected to the SmartGuard/SeaGuard Datalogger is
shown. This makes it possible to get a correct conductivity input
from another sensor through the datalogger
7)
Only in AiCaP
mode
Conductivity
Parameter Id
ENUM 1
A list of all available Conductivity parameters with mS/cm as
unit is shown in the dropdown menu. Make sure that the correct
conductivity parameter is selected. Only in AiCaP mode.
Heading Sensor Id
ENUM 1
When the DCPS is connected to AiCaP, a list of other available
sensors connected to the SmartGuard/SeaGuard Datalogger is
shown. This makes it possible to get a correct heading input
from another sensor through the datalogger
7)
Only in AiCaP
mode