July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
Page 25
C1 Enable
Surface
Reference
BOOL 1
Enables surface reference for Column 1. If not selected the column will
be instrument referenced.
Note! The sensor needs a pressure input to
compensate for tidal variations. Cannot be used it in combination with
downward orientation.
SC No /
Low
C1 Cell Size
ENUM 1
The Cell size setting for Column 1 in meter. Selectable from 0.5 to 5
meter. Default setting is 1 meter.
C1 Distance
first Cell
Center
ENUM 1
The distance to centre of first cell for Column 1 in meter, distance from
surface if surface referenced or sensor if not surface referenced.
Always sensor referenced if used upside down. Selectable from 1 to 70
meter. Default setting is 3 meter.
C1 Number Of
Cells
ENUM 1 Number of Cells in Column 1, maximum 75 cells. Default setting is 45.
C1 Cell Center
Spacing[m]
ENUM 1
Spacing between two cells for Column 1 in meter. If spacing is equal to
cell size the two cells is next to each other. Spacing less than cell size
gives overlap and spacing more than cell size gives space between two
cells. Selectable from 0.1 to 30 meter. Default setting is 1 meter.
C2 Enable
Column
BOOL 1 Enables Column 2 (Column 1 is always enabled)
C2 Enable
Surface
Reference
BOOL 1
Enables surface reference for Column 2. If not selected the column will
be instrument referenced.
Note! The sensor needs a pressure input to
compensate for tidal variations. Cannot be used it in combination with
downward orientation.
C2 Cell Size
ENUM 1
The Cell size setting for Column 2 in meter. Selectable from 0.5 to 5
meter. Default setting is 1 meter.
C2 Distance
first Cell
Center
ENUM 1
The distance to centre of first cell for Column 2 in meter, distance from
surface if surface referenced or sensor if not surface referenced.
Always sensor referenced if used upside down. Selectable from 1 to 70
meter. Default setting is 3 meter.
C2 Number Of
Cells
ENUM 1 Number of Cells in Column 2, maximum 50 cells. Default setting is 10.
C2 Cell Center
Spacing[m]
ENUM 1
Spacing between two cells for Column 2 in meter. If spacing is equal to
cell size the two cells is next to each other. Spacing less than cell size
gives overlap and spacing more than cell size gives space between two
cells. Selectable from 0.1 to 30 meter. Default setting is 2 meter.
C3 Enable
Column
BOOL 1 Enables Column 3
C3 Enable
Surface
Reference
BOOL 1
Enables surface reference for Column 3. If not selected the column will
be instrument referenced.
Note! The sensor needs a pressure input to
compensate for tidal variations. Cannot be used it in combination with
downward orientation.
C3 Cell Size
ENUM 1
The Cell size setting in Column 3 in meter. Selectable from 0.5 to 5
meter. Default setting is 1 meter.