English      2 

1.

 

Dear Customers 

Congratulations on purchasing a state-of-the-art A&D sphygmomanometer, 
one  of  the  most  advanced  monitors  available  on  the  market  today. 
Designed  for  ease  of  use  and  accuracy.  We  recommend  that  you  read 
through this manual carefully before using this device for the first time. 

 

Intended User:

   

This device is designed to be used by people 
who can measure blood pressure by 
auscultation. 

Applicable Patient:

   

The device is designed for use on adults and 
children aged 3 and older. 

 

Environment for Use:

    The device is designed for use in a medical 

facility. 

 

Intended purpose:

   

This device is designed to measure human 
blood pressure and heart rates. 

 

Features 

Measurement 

  This sphygmomanometer is designed to monitor and display cuff pressure 

during cuff inflation (by rubber bulb) and deflation (by exhaust bulb) while 
the  user  determines  the  patient’s  blood  pressure  level  by  listening  for 
Korotkoff sounds with a stethoscope. 

 

Safety 

  This sphygmomanometer is designed to measure a patient’s blood pressure 

without the use of mercury, therefore protecting your local environment. 

  An automatic quick exhaust valve is installed in the device to prevent over 

pressurization of 300mmHg or higher, therefore protecting the patient. 

 
 

Summary of Contents for UM-102

Page 1: ...Instruction Manual Original Manuel d instructions Traduction Manual de instrucciones Traducci n Manuale di Istruzioni Traduzione 1WMPD4002909B...

Page 2: ......

Page 3: ...ction Manual Contents 1 Dear Customers 2 2 Preliminary Remarks 3 3 Notes for Proper Use 4 4 Parts Identification 7 5 Symbols 8 6 Using the Monitor 9 7 Measurements 12 8 Useful Features 13 9 Troublesho...

Page 4: ...se The device is designed for use in a medical facility Intended purpose This device is designed to measure human blood pressure and heart rates Features Measurement This sphygmomanometer is designed...

Page 5: ...parately if the device will not be used for a month or more Otherwise the batteries may degrade The battery life varies with the ambient temperature and may be shorter at low temperatures Generally tw...

Page 6: ...rated oxygen environments such as a high pressure oxygen chamber or an oxygen tent It may cause a fire or explosion When the cuff is infected by blood or body fluid it should be safely disposed of acc...

Page 7: ...through the artery This may cause pain numbness or a temporary red mark to the arm This condition will appear especially when measurement is repeated successively Any pain numbness or red marks will...

Page 8: ...ter exercising or taking a bath A patient must rest for twenty to thirty minutes before the measurement Notes for Proper Measurement Let a patient sit down in a comfortable position Confirm that the p...

Page 9: ...merical display During measurement After exhausted air Indicates pressure value Indicates pulse rate Display Unit of pressure Pressure value Battery indicator Pulse rate Unit of pulse High digits of t...

Page 10: ...s Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Pressure remains in the cuff E...

Page 11: ...ries It may shorten the battery life or cause the device to malfunction Battery life varies with the ambient temperature and may be shorter at low temperatures Only the specified batteries should be u...

Page 12: ...cuff CUF KW LL 31 cm to 45 cm Large Adult cuff CUF KW LA 22 cm to 32 cm Adult cuff CUF KW A 16 cm to 24 cm Small Adult cuff CUF KW SA 12 cm to 17 cm SS cuff CUF KW SS Arm size The circumference of th...

Page 13: ...nge ARTERY Artery position mark THIS SIDE TO PATIENT Instructions to the patient REF Catalog number Lot number ADULT Cuff size 22 cm to 32 cm LL Cuff size 41 cm to 50 cm LARGE ADULT Cuff size 31 cm to...

Page 14: ...ial artery and pressurize the cuff by squeezing the rubber bulb Make sure the exhaust valve is completely closed While the cuff is inflating the pressure bar will move and in turn the LCD will display...

Page 15: ...ction Allow at least three minutes between measurements on the same person 8 Useful Features Measurement with MARK Button You can put a marker at a certain pressure value when the MARK button is press...

Page 16: ...tact your nearest authorized A D dealer Do not attempt to open or repair this product by yourself as any attempt to do so will render your warranty invalid 10 Maintenance Do not attempt to open the de...

Page 17: ...enzalkonium Chloride Solution 0 05 Sodium Hypochlorite 0 05 Do not wipe off the valve and its vicinity with gauze soaked in disinfectant solution It will cause stuffing of the rubber bulb Clean the de...

Page 18: ...to 40 C 15 RH to 85 RH 800 hPa to 1060 hPa Transport Storage conditions 20 C to 60 C 10 RH to 95 RH 700 hPa to 1060 hPa Dimensions Approx 98 W x 324 H x 67 D mm Weight Approx 520 g excluding batteries...

Page 19: ...10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Proximity fields from RF wireless communications equipment IEC 61000 4 3 See table 3 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Tabl...

Page 20: ...English 18 MEMO...

Page 21: ...lifornia 95131 U S A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D RUSO...

Reviews: