English 15
Cleaning
When cleaning the device and cuff, turn off the power by pressing
START/STOP
button and remove the batteries.
When the main body or cuff is dirty, wipe them fully by using a gauze or
cloth dampened with water or warm water avoiding excess water.
To prevent a risk due to infection, disinfect the main body and cuff regularly.
When disinfecting them, wipe them gently by using the gauze or
dampened cloth with local antiseptic solution then wipe the moisture off the
surface by using a dry soft cloth.
Use the following disinfectants to clean the main body and cuff.
Ethanol (70%)
Isopropanol (70%)
Chlorhexidine Gluconate Solution (0.5%)
Benzalkonium Chloride Solution (0.05%)
Sodium Hypochlorite (0.05%)
Do not wipe off the valve and its vicinity with
gauze soaked in disinfectant solution. It will
cause stuffing of the rubber bulb.
Clean the device about once every month, based on the policy or
instruction specified by the user authority.
Disposal
This unit and its batteries are not treated as ordinary household waste and
must be disposed of according to the applicable local regulations.
Item
Parts
Material
Package
Box
Cardboard
Cushion
Cardboard
Bag
PE
Main unit and
accessories
Enclosure
ABS
Internal parts
General electronic components
Batteries
Alkaline battery
11.
Technical Data
Model
UM-102
Measurement method
By applying pressure, enabling the determination
of blood pressure by the auscultatory method.
Measurement range
Numerical display
Pressure:
0 - 300 mmHg
Rubber bulb
Valve