64
Code Meaning
Cause
and
treatment
Out of range,
30
≦
PUL
≦
200
Out of range,
30
≦
DIA
≦
160
Out of range,
60
≦
SYS
≦
280
Out of range,
10
≦
PP
≦
150
#1
If these errors occur multiple times, try
another blood pressure measurement.
#1
PP
=
SYS
-
DIA
SYS :
Sys
tolic blood pressure
DIA :
Dia
stolic blood pressure
PP :
P
ulse
p
ressure
Measurement is
above 180 seconds.
If the inflation speed or exhaust speed is
slow, an inspection is necessary.
Exhaust is above 90
seconds.
The exhaust speed may be slow, an
inspection is necessary.
Heartbeat cannot be
detected.
Heartbeat cannot be detected
because of body motion, etc.
Measure the blood pressure while
relaxed and do not moving.
The settings of the
interval time are
incorrect.
If the interval time is set to 120 minutes,
the difference between last
start time
and next
start time
cannot divide into
two hours perfectly.
Zero pressure error
for safety circuit.
Displays at the measurement start
time.
Release the air remaining in the cuff
completely.