63
9.6.
Error Codes
Measurement error codes
Code Meaning
Cause
and
treatment
Pressure zero error
Release the air left in the cuff.
Low battery
Replace with new batteries.
Failure of
pressurization
Inflation does not reach the target
pressure.
Confirm the cuff connection.
If there are no problems with the cuff
connection, the recorder may have
malfunctioned and requires
inspection.
Pressure exceeds
299 mmHg
Body motion may be occurred in the
pressurization. Relax and keep the
silence during measurement. If the
treatment is not enough, inspect the
recorder.
Force stop using
START/STOP
switch.
Press the START/STOP switch only
when necessary.
Blood pressure cannot
be measured.
The heartbeat cannot be detected due
to body motion or noise from clothes.
Relax and do not move.
Confirm the position of the cuff.
If this failure occurs even when
relaxed, contact your dealer to
inspect and repair the recorder.
Excessive body
motion.
Relax and keep silent during the
measurement.