22
6.
Blood pressure measurement Functions
The recorder is equipped with automatic blood pressure measurement
(A
-
BPM) and can store measurement states and measurement
results.
6.1.
Automatic Blood Pressure Measurement (A-BPM)
When the A-BPM function is not used, suspend the function
by pressing and holding the EVENT switch for 3 seconds
or more so that the
mark turns off. Otherwise, the
measurement will start at the next start time and the cuff may
burst.
The
A
-
BPM function measures the blood pressure at preset
intervals using the built-in clock and stores the measurement
result in the memory.
A
-
BPM can be started and suspended by pressing and holding
the EVENT switch 3 seconds or more.
The mark is displayed on the OLED display while A
-
BPM is used.
Blood pressure is measured automatically at the A
-
BPM start time.
The initial pressurization value is set to
180
mmHg at the factory.
If the first pressurization is not enough, re-pressurizations are
performed automatically up to two times.
When you delete data in the memory or suspend A-BPM, the
pressurization value is reset to the initial pressurization value.
When a measurement error occurs and the waiting time until the
Summary of Contents for TM-2440
Page 18: ......
Page 32: ...14 4 2 Dimensions 95 Unit mm 24 5 ...
Page 94: ...76 MEMO ...
Page 95: ......