45
How to stop or suspend the measurement
Press the START/STOP switch to stop blood pressure
measurement. An error code is stored in the memory. Blood
pressure is measured again after
120
seconds.
Concerning of A-BPM, only the current blood pressure
measurement is stopped, and measurement will be performed at
the next
start time
.
To suspend A-BPM, press and hold the EVENT switch for 3
seconds or more so that the
mark turns off.
Remove the cuff if the current blood pressure measurement can
not be stopped using the START/STOP switch.
Press the START/STOP switch to stop blood pressure
measurement. An error code is stored in the memory.
During A-BPM, only the current blood pressure
measurement is stopped, and measurement will be
performed at the next
start time
.
When a pain of the arm or unexpected condition occur, stop
the measurement, remove the cuff and consult the doctor.
Suspend A-BPM by pressing and holding the EVENT
switch for 3 seconds or more so that the
mark turns off.
Press and hold the EVENT switch for 3 seconds or more
again to resume A
-
BPM automated measurement. The
mark is
shown on the OLED display. Recording of data is carried out
continuously except during the suspended period.