Why may the MS/MX/MF/ML
have a lower moisture content
rate and longer measurement
time than the FD-620 with the
same sample?
1. The FD-620 displays a measurement temperature lower
than the actual temperature above the sample. Therefore,
the MS/MX/MF/ML heats to a higher temperature and tends
to have longer measurement times and lower content
values than the FD-620.
2. The MS/MX/MF/ML improves on the FD-620 and was
designed so the actual temperature above the pan matches
the setting and display temperatures. If you want to get the
same result as the FD-620, reducing the sample mass by
1g and setting the end point condition to HI should produce
a similar result.
No.
Question
Answer
5
What is highly accurate SHS?
SHS stands for Super Hybrid Sensor. A&D created this weight
sensor to enable weighing within 1 second on average and
significantly improve accuracy and long-term stability. This
sensor is used in scales that require high resolution and
accuracy.
The SHS enables the MS/MX/MF/ML moisture analyzer to
capture dynamic weight changes during heating. Furthermore,
the high resolution and accuracy allow for smaller sample
sizes. These benefits enable the moisture content to be
measured in shorter length of time. (Patent approved)
6
Why are two pan handles
included?
If a pan is used for consecutive measurements and a sample is
placed on a still-hot pan, water may evaporate before the start
of measurement and make the measurement data inaccurate.
To avoid this, alternate two pans and handles. (The ML has one
handle only because it has a lower resolution of 0.1%.)
Alternating enhances the reliability and repeatability of results.
Furthermore, it improves operability during repeated
measurements and prevents mishaps like burns.
■
Introduction to moisture analyzers/B. Measurement