HVG-WP Series
Page 43
14.2.3.
Calibration of the Zero Point
Step 8 After 30-minute warm up, press and hold
the
CAL
key to enter the calibration mode.
Then
Cal 0
is displayed
Step 9 Confirm that nothing is placed on the pan.
Wait for the stability mark to be displayed.
Press the
ENTER
key. The scale stores
the current condition as the zero point.
Step 10 The scale displays
5pn1
for several
seconds. Calibration of zero point is
finished. To exit the calibration mode,
proceed to step 14.
14.2.4.
Span Calibration
Step 11 Set the value of the calibration mass
using the following keys (This initial value
depends on the scale mode).
∧
key
Selecting the figure to be
changeed.
<
key
Change the number of the
figure.
Step 12 Place the mass on the pan which was set
at step 11. Wait for the stability mark to be
displayed. Press the
ENTER
key. The
scale then calculates the span and stores
it.
Step 13 The scale displays
End
at the finish.
Remove the mass from the weighing pan.
Step 14 Press the
CAL
key to return to the normal
weighing mode.