Page 38
HVG-WP Series
Caution
The parameters of the upper limit value (HI) and a final value (HI) use the
same memory. The parameters of the lower limit value (LO) and the zero band
(LO) use the same memory.
The upper/lower comparator function, the simple batch function and the
full/dribble batch function can not be used at the same time because these
parameters use common memory.
13.1.
Preparation (Setting Parameters)
Step 1 Turn off the display.
Press
the
ON/OFF
key while the
ZERO
key is pressed
and held. F1 is displayed.
Step 2 Press the
ENTER
key repeatedly to display F6.
Step 3 Select a parameter of 8 for the simple batch function
with the
∧
key.
Step 4 Press the
ENTER
key to store the new parameter.
Step 5 Press the
F
key and the
ENTER
key to exit from the
function table. The scale returns to the weighing
mode.