HVG-WP Series
Page 33
12.
Full/Dribble Batch Function
This function allows the scale to control a 2 speed batching process with full and
dribble feed conditions.
To use this funtion, the following parameters need to be set: F6 to setting 9, F10 to the
setting of your choice, HI(dribble), OK(full feed) and LO(zero band).
Install option OP-03 or OP-04, to use the relay output of the full/dribble batch function.
The functionality and performance of this function are as follows:
Zero band
indicator/
LO relay output
Preliminary value
indicator/
OK relay output
Final value
indicator/
HI relay output
A
A
A
Gross
<
Zero band
(Zero detection Level)
ON / Make
OFF / Break
OFF / Break
A
Net
<
Final - Preliminary
OFF / Break
ON / Make
ON / Make
A
Final - Preliminary
≦
Net
<
Final
OFF / Break
OFF / Break
ON / Make
A
Final
≦
Net
OFF / Break
OFF / Break
OFF / Break