HVG-WP Series
Page 9
5.1.
Removing
the
column
Caution
Remove the main power plug before removing the column (type V)
Remove the AC adapter before removing the column (Type L)
When removing the loadcell cable, do not pull the loadcell cable connector
forcibly and do not pull on the wires of the cable.
Do not bend the cable forcibly.
Avoid dust, static electricity and high humidity (or drops) because the inside
of the indicator unit is very sensitive.
Procedure
Step 1 Remove the power plug from the power outlet.
(Type V). Remove AC adapter from the
Power Jack (Type L).
Step 2 Open the rear cover of indicator unit.
Disconnect the loadcell cable connector gently
(perpendicularly, do not pull to the side).
Step 3 Remove the ferrite core and cable clamp from
the loadcell cable.
Step 4 Loosen the knobs to remove the indicator unit.
Step 5 Remove four 6mm screws from the bottom of the
column.
Step 6 Carefully remove the cable from the pole and
pole base so that the connector is not pulled
forcibly.
Step 7 Arrange the cable so that it does not touch the
weighing pan in the base unit. The untied cable
is at least 2m long. The optional extension
loadcell cable (OP-02) is 5m long.
knob
Loadcell cable
Cable clamp
Ferrite core
Wind up 2 times
Screw and O ring
Connector
Pole