GX-A Series/GF-A Series Communication Manual 38 / 41
This is a list of items related to communication of internal setting values. For other items, refer to
the GX-A / GF-A instruction manual.
Classification
item
Setting item
Setting
value
Contents, usage
Refer to the GX-A / GF-A instruction manual
■
Key mode
Data output with the PRINT
key when the weighing value
is stable
Auto print A mode
(
reference = zero point
)
Data output of a stable
weighing value when it
exceeds the range of
and in relation to
zero.
Auto print B mode
(
reference = previous
stable value
)
Data output of a stable
weighing value when it
exceeds the range of
and in relation to
previous stable value.
Strem
mode Output every time when the
display value is renewed
Key mode B mode
(
immediate output
)
Data output regardless of
stability / instability with the
PRINT key
Key mode C mode
Data is output when stable
with the PRINT key and if
unstable, it is output once is
has stabilized
Data output mode
Interval
mode Data output after every cycle
set by the setting
Only
plus
When greater than reference
value
Only
minus
When smaller than reference
value
Auto print Polarity
■
Bipolarity
Regardless of the value size
compared with reference value
■
10 digits
100
digits
Data output
Auto print difference
1000
digits
Difference from reference
value
□
“
■
” Factory setting.
□
”1 digit” stands for a smallest displayed order. In case of GX-303A 1 digit is 0.001g.