GX-A Series/GF-A Series Communication Manual 14 / 41
As for the balance’s data output timing, it can be changed with the internal setting (data
output mode).
Key mode
Internal setting
If the PRINT key is pressed when the stable value mark is displayed, the weighing value will be
output once.
At that time the displayed weighing value will blink once to indicate that it had been output.
Auto print A mode
Internal setting
When the weighing value exceeds the range specified by the internal setting (auto print
polarity) and the internal setting (auto print width) from the standard "zero display" and the
stable value mark is on, the weighing value will be output once. Also, if the PRINT key is pressed
while the stable value mark is on, the weighing value will be output once.
At that time the displayed weighing value will blink once to indicate that it had been output.
Related internal settings
Auto
polarity
Auto
width
Auto print B mode Internal setting
When the weighing value exceeds the range specified by the internal setting (auto print
polarity) and the internal setting (auto print width) from the standard "value previously
displayed with a stable value mark" and the stable value mark is on, the weighing value will be
output once. Also, if the PRINT key is pressed while the stable value mark is on, the weighing value
will be output once. At that time the displayed weighing value will blink once to indicate that it had
been output.
Related internal settings
Auto
polarity
Auto
width
Stream mode
Internal setting
Regardless of the presence or absence of the stable value mark, weighing value will be output for
each internal setting (display rewrite cycle). When the internal setting is (5 times /
sec), the output is at approximately 5.21 Hz.
Related internal settings
Display
rewrite
cycle
Baud rate
Caution
□
Depending on the display rewrite cycle and the baud rate, all data may not be transmitted
unless the baud rate is increased.
7. Data output
7-1. Data output mode