18
Each of the four Summary Data is titled with the name of the tab, file, patient and measurement
method. For detailed information on Summary Data screen listings, see below:
Summary Data Tab
Definition
Time Interval
The analysis period, start time/date – end time/date, for the
activated Summary Data tab.
Valid Readings
The total number of measurements taken during the monitoring
session less any monitoring errors or manually excluded readings
within the data set. See “
” for
information on measurement errors.
Excluded Readings
The measurement error percentage. This figure is based on the
combined total of measurement errors and excluded
measurements.
Min (Minimum)
The minimum measurement value for Systolic, Diastolic, MAP
and Pulse measurements during the Time Interval.
Mean
The mean value for Systolic, Diastolic, MAP and Pulse
measurements during the Time Interval.
Max (Maximum)
The maximum measurement value for Systolic, Diastolic, MAP
and Pulse measurements during the Time Interval.
SD (Standard Deviation)
The Standard Deviation of the value for Systolic, Diastolic, MAP
and Pulse measurements during the Time Interval.
SE (Standard Error)
The Standard Error of the value for Systolic, Diastolic, MAP and
Pulse measurements during the Time Interval.
CV(%)
(Coefficient of Variation )
The Coefficient of Variation for Systolic, Diastolic, MAP and Pulse
measurements during the Time Interval.
SYS/DIA limits
Systolic and Diastolic limits for the Awake and Sleep interval.
Readings > SYS limit(s)
The percentage of the Systolic measurement values greater than
the specified Systolic limit. See “
" for information on specifying the Systolic Blood
Pressure Limit.
Readings > DIA limit(s)
The percentage of the Diastolic measurement values greater
than the specified Diastolic limit. See "
" for information on specifying the Diastolic
Blood Pressure Limit.
Highest Systolic Value
The time and date of the highest Systolic value during the Time
Interval.
Highest Diastolic Value
The time and date of the highest Diastolic value during the Time
Interval.
Lowest Systolic Value
The time and date of the lowest Systolic value during the Time
Interval.
Lowest Diastolic Value
The time and date of the lowest Diastolic value during the Time
Interval.
Summary of Contents for Doctor Pro TM-2430-13
Page 54: ...53 4 PRINTOUT EXAMPLES Figure 36 Patient information Text data ...
Page 55: ...54 Figure 37 Summary data Full and partial Text data ...
Page 56: ...55 Figure 38 Summary data Awake and sleep Text data ...
Page 57: ...56 Figure 39 Blood pressure data Text data ...
Page 58: ...57 Figure 40 Trend graph Blood pressure and double product Graphical data ...
Page 59: ...58 Figure 41 Correlation plots upper and histogram plots lower Graphical data ...
Page 60: ...59 Figure 42 Circadian rhythm plots upper and HBI analysis plots lower Graphical data ...
Page 61: ...60 Figure 43 Mini report 1 ...
Page 62: ...61 Figure 44 Mini report 2 ...
Page 67: ...66 MEMO ...
Page 68: ...67 MEMO ...
Page 69: ...68 MEMO ...