11.2.
GLP Report
Set the following parameters to output the GLP / GMP compliant report.
Refer
to
"
16.2. Connection To Peripheral Equipment
" for connection to an AD-8121B.
To print the report, set the "GLP output (
info
)" parameter to "
1
" and use MODE 3 of the AD-8121B. If
MODE1 is used, select temporary dump print mode by pressing the
STAT.
key of the AD-8121B.
To output the report to a personal computer using the RS-232C interface, set the "GLP output (
info
)"
parameter to "
2
".
If the time and date are not correct set the correct time and date in "Clock (
Cl
adj
)" of the function table.
Calibration report using the internal mass
Key operation
1 Press
the
CAL
key to display
Cal
in
and calibrate the balance automatically.
2 If GLP output is used,
glp
is displayed and the calibration report is output.
3 The balance returns to weighing mode after this calibration.
Setting
of
"
info
1
" Setting
of
"
info
2
"
AD-8121 printer format General
format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage
return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
===========
A
=
&
=
D
MODEL
=====
BM-300
S/N
=====
01234567
ID
======
LAB-0123
DATE
==
2011/02/20
TIME
====
12:34:56
CALIBRATED(INT.)
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-