![A&D BM-20 Manual Download Page 73](http://html1.mh-extra.com/html/aandd/bm-20/bm-20_manual_3108747073.webp)
BM series
71
1 2
3 4
MODE
ON
1 2
3 4
MODE
ON
16.2.
Connection To Peripheral Equipment
Connection to an AD-8121B printer
Preset the following parameters to use the AD-8121B printer.
Class Item
and
Parameter
Factory
settings
AD-8121B
MODE 1
AD-8121B
MODE 2
AD-8121B
MODE 3
prt
Data output mode
0
0
,
1
,
2
3
0
,
1
,
2
ap-p
Auto print polarity
0
ap-b
Auto print difference
1
#1
Not
necessary
#1
5-id
ID number output
0
off
off
off
,
on
d-no
Data number output
0
off
off
off
,
on
5-td
Time/Date output
0
0
0
0
,
1
,
2
,
3
pU5e
Data output pause
0
off
off
off
,
on
#2
dout
Data output
at-f
Auto feed
0
off
off
off
,
on
bp5
Baud rate
2
2
2
2
btpr
Data bit, parity bit
0
0
0
0
Crlf
Terminator
0
0
0
0
5if
Serial
interface
type
Data format
0
0
0
1
#1
Set parameters when auto print mode A or B (
prt
1
or
2
) is selected.
#2
Set
1
when multiple lines are printed. Example: When appending ID number, set
1
.
Notes
Refer
to
"
11.2. GLP Report
" concerning print samples.
Settings
of
AD-8121B
MODE
AD-8121B
DIP switch
Description
MODE 1
1 2
3 4
MODE
ON
Print at receiving data.
Standard mode, statistic mode
MODE 2
1 2
3 4
MODE
ON
Print by DATA key operation or built-in timer.
Standard mode, interval mode, chart mode
MODE 3
1 2
3 4
MODE
ON
Print at receiving data.
Dump print mode
DIP switch No.3 : Handling unstable data
ON
OFF
Not
printed
DIP switch No.4 : Data input specifications (Interface selection)
ON
Current
loop
OFF
RS-232C
Summary of Contents for BM-20